Details
6-Series sondes are going away. Read about the discontinuation at YSI.com/6-series
Designed for use in fresh, sea or polluted waters, the 600OMS V2 utilizes the field-proven YSI sensors and incorporates innovations in sensor configuration such as a conductivity and temperature module that fits into the sonde body.
Parameters available for the 600OMS V2 include:
- Temperature
- Conductivity
- Salinity
- Specific Conductance
- Depth
- TDS
- Shallow Vented Level
- Optical Dissolved Oxygen, Blue-green Algae, Turbidity, Chlorophyll or Rhodamine
The 600OMS V2 is available with or without internal power. Its small size is perfect for applications such as turbidity or oxygen monitoring.
Features:
- Fits down 2-inch wells
- Easily connects to data collection platforms
- Compatible with the YSI 650 MDS
- Wiped optics prevent biofouling
Specification:
Certifications | CE, FCC, Assembled in the USA, WEEE, and MCERTS, IP-67, EU Battery Compliance |
---|---|
Connectivity / Communications | RS-232, SDI-12 |
Desktop Software Compatible | EcoWatch® |
Medium | Fresh, sea or polluted water |
Monitoring | Yes |
Multiparameter | Yes |
Operating Temperature | -5 to +50°C |
Parameters Measured | temperature/conductivity, dissolved oxygen, turbidity, chlorophyll, blue-green algae, rhodamine |
Power | 25 to 30 days at 15 minute sampling interval at 25°C, 4 AA Alkaline cells |
Sampling | Yes |
Storage Temperature | -10 to +60°C |
User Calibratable | Yes |
Waterproof | IP-67 |
General Sonde Specifications | |
---|---|
Medium | Fresh, sea or polluted water |
Temperature - Operating | -5 to +50°C |
Temperature - Storage | -10 to +60°C |
Communications | RS-232, SDI-12 |
Software | EcoWatch® |
Diameter | 1.65 in, 4.2 cm |
Length | 21.3 in, 54.1 cm |
Weight | 1.3 lbs, 0.6 kg |
Weight with Batteries | 1.4 lbs, 0.7 kg |
Power - External | 12 V DC |
Power - Internal (optional) | 4 AA Alkaline cells, 25 to 30 days at 15 minute sampling interval at 25°C |
Certifications | CE, EU Battery Compliance, FCC, IP-67, WEEE, and MCERTS; Assembled in the USA |
Sensor Specifications | Range | Resolution | Accuracy |
---|---|---|---|
ROX™ Optical Dissolved Oxygen• % Saturation |
0 to 500% | 0.1% | 0 to 200%: ±1% of reading or 1% air saturation, whichever is greater; 200 to 500%: ±15% of reading |
ROX™ Optical Dissolved Oxygen• mg/L |
0 to 50 mg/L | 0.01 mg/L | 0 to 20 mg/L: ± 0.1 mg/L or 1% of reading, whichever is greater; 20 to 50 mg/L: ±15% of reading |
Conductivity•• | 0 to 100 mS/cm | 0.001 to 0.1 mS/cm (range dependent) |
±0.5% of reading + 0.001 mS/cm |
Salinity | 0 to 70 ppt | 0.01 ppt | ±1% of reading or 0.1 ppt, whichever is greater |
Temperature | -5 to +50°C | 0.01°C | ±0.15°C |
Depth - Medium | 0 to 200 ft, 61 m | 0.001 ft, 0.001 m | ±0.4 ft, ±0.12 m |
Depth - Shallow | 0 to 30 ft, 9.1 m | 0.001 ft, 0.001 m | ±0.06 ft, ±0.02 m |
Vented Level | 0 to 30 ft, 9.1 m | 0.001 ft, 0.001 m | ±0.01 ft, 0.003 m |
Turbidity• 6136 Sensor‡ |
0 to 1,000 NTU | 0.1 NTU | ±2% of reading or 0.3 NTU, whichever is greater** |
Rhodamine• | 0-200 μg/L | 0.1 μg/L | ±5% reading or 1 μg/L, whichever is greater |
Blue-Green Algae Sensor Specifications | Range | Detection Limit | Resolution | Linearity |
---|---|---|---|---|
Blue-Green Algae Phycocyanin• |
~0 to 280,000 cells/mL† 0 to 100 RFU |
~220 cells/mL§ | 1 cell/mL 0.1 RFU |
R2 > 0.9999** |
Blue-Green Algae Phycoerythrin• |
~0 to 200,000 cells/mL† 0 to 100 RFU |
~450 cells/mL§§ | 1 cell/mL 0.1 RFU |
R2 > 0.9999*** |
Chlorophyll• 6025 Sensor‡ |
~0 to 400 μg/L 0 to 100 RFU |
~0.1 μg/L§§§ | 0.1 μg/L Chl 0.1% RFU |
R2 > 0.9999**** |
View more about YSI 600OMS V2 Optical Monitoring Sonde on main site